SITRANS TH300 Temperature Transmitter
Brand: Siemens AGSITRANS TH300 with HART protocol is designed to support all common thermocouples, resistance and millivolt sensors.
Setup is quick and easy with SIMATIC PDM or a handheld communicator.
Unique user-friendly details are implemented, one of them being the direct readout of the output current with a multimeter without having to open the 4 to 20 mA loop.
A red/green LED indicates the device status at a glance.
The SITRANS TH300 is configured over HART.
This can be done using a handheld communicator or even more conveniently with a HART modem and the SIMATIC PDM parameterization software.
The configuration data are then permanently stored in the non-volatile memory (EEPROM).
Once the sensors and power supply have been correctly connected, the transmitter outputs a temperature-linear output signal and the diagnostics LED displays a green light.
In the case of a sensor break, the LED flashes red, an internal device fault is indicated by a steady red light.
The test socket can be used to connect an ammeter at any time for monitoring purposes and plausibility checks.
The output current can be read without any interruption, or even without opening the current loop.
Features- Measures temperature with a linear output signal
- High accuracy across entire ambient temperature range
- Reduces plant noise
- Alarm signal for sensor break or short-circuit according to NAMUR 43
- Electro magnetic compatibility (EMC) according to DIN EN 61326 and NE21
- Save expensive compensation cables with internal reference junction
- Galvanic insulation for accuracy and safety in thermocouple applications
- Single- or dual-point trim
- Custom characteristic curve for application of non-standard sensors
- Difference- and average measurement
- Extensive diagnosis: e.g. drag indicator and time meter
- Set of output current for simulation
- Explosion protected variants for US, Canada and Europe
- Rugged design, fully potted electronics
Specifications
Brand
Output Signal
- 4-20 mA, 2-Wire
Auxiliary Power
- 11-35 Vdc
Load
- 500 ohms
Error Due to Internal Reference Junction
- < 0.5° C
Ambient Temperature
- -40° C to 85° C
- -40° F to 1855° F
Electromagnetic Compatibility
- acc. to EN 61326 and NE21
Relative Humidity
- < 98 %, with condensation
Material
- Molded Plastic
Weight
- 50 g
Degrees of Protection
- Enclosure: IP40
- Terminals: IP00
Certifications & Approvals
- PTB 05 ATEX 2040X
- II 1 G Ex ia IIC T6/T4
- II 2 (1) G Ex ia/ib IIC T6/T4
- II 3(1) G Ex ia/ic IIC T6/T4
- II 1D Ex iaD 20 T115° C
- II 3 G Ex nL IIC T6/T4
- II 3 G Ex nA IIC T6/T4
- FM 3024169
- FM 3024169C
- IS / Class I, II, III / Div. 1 / GP ABCDEFG T6, T5, T4
- EAC Ex(GOST)
- NEPSI
- IEC
- EXPO-LABS
Units of Measure
- °C
- °F
Sensor Current
- < 0.45 mA
Response Time
- < 250 milliseconds for 1 sensor with break monitoring
Sensor Type
- Pt100 (IEC) 2-Wire
- Pt100 (IEC) 3-Wire
- Pt100 (IEC) 4-Wire
- Thermocouple Type: B, C, D, E, J, K, L, N, R, S, T, U
- SITRANS TH300 transmitters can be used in all industrial sectors.